INTAKE MANIFOLD > INSPECTION |
| 1. INSPECT INTAKE MANIFOLD |
![]() |
![]() |
Disconnect the vacuum hose from the actuator.
Check the operation of the swirl control valve.
Check that the valve is fully opened under normal conditions.
Connect a vacuum pump to the control valve. Check that the valve is fully closed when a negative pressure of 35 kPa (270 mmHg, 10 in.Hg) is applied.
Leave it for 1 minute and check that the vacuum pump needle does not move.
If the result is not as specified, replace the intake manifold.
| 2. INSPECT VACUUM SWITCHING VALVE (for SWIRL CONTROL VALVE) |
![]() |
Measure the resistance of the VSV.
| Tester Connection | Specified Condition |
| 1 - 2 | 33 to 39 Ω at 20°C (68°F) |
| 1 - Body ground 2 - Body ground | 1 MΩ or higher |
![]() |
Check the operation of the VSV.
Check that air flows from port E to the filter.
If the result is not as specified, replace the VSV assembly.
![]() |
Apply battery voltage across the terminals.
Check that air flows from port E to port F.
If the result is not as specified, replace the VSV assembly.